SPECTRAL IMAGING: Real-time spectral unmixing enhances understanding of cellular biomarkers

By Murray Reed, Elliot Wachman, Austin Dowd, and Alex Fong

The use of multiple molecular biomarkers theoretically enables both location detail and clear interpretation of tissue abnormalities, but in practice can complicate analysis. Aided by acousto-optic tunable filters and spectral unmixing, however, multispectral imaging cuts through the complications, promising benefits for applications from vital signs monitoring and cancer detection to neuroscience.

Microscopic imaging of cellular sample morphology has been used to diagnose medical disorders for centuries. Recent years have brought the discovery of numerous molecular biomarkers for various diseases—and now, simple chemical assays are playing an increasing role in primary diagnosis and determination of medical treatment. But because individual biomarkers generally lack the sensitivity and specificity to completely identify a disease state, researchers yearn for the simultaneous use of multiple biomarkers.

Combining multiple molecular biomarkers with traditional microscopy would mean that biomarkers could be used both to locate potential abnormal regions on a microscope slide and to clarify the interpretation of those abnormalities. But coupling biomarkers to stains to make them easily recognizable often confuses things, because the colors become hard to interpret by eye.

Analyzing such complex microscope slides with multispectral imaging, however, provides a detailed spectrum for every pixel in the image. This article describes a multispectral system optimized for fast and efficient analysis of traditionally stained clinical pathology samples to which molecular labels have been added. The instrument can clearly display results in real time as the magnified image is scanned across a large slide.

Multispectral brightfield microscopy

Multispectral imaging systems provide detailed spectral information for every pixel. If the pixels are imagined as arranged in a matrix in the x-y plane, the information can be displayed along the z-axis in a 3D data structure called an "image cube." Linear spectral unmixing is a spectral imaging analysis technique: It assumes that the total acquired spectra at each pixel can be expressed as a linear combination of contributions from a series of independent elements, each having its own distinct "reference spectrum." The unmixing technique allows the final image to be decomposed into separate "rule images," one for each of the reference spectra. The intensity of each rule image corresponds to the calculated amount of that spectral component present at each pixel. If the image of a microscope slide contains overlapping spectra from multiple biomarkers that are difficult to distinguish visually, multispectral imaging makes it possible to analyze the image and spectrally unmix the reference spectra of each individual biomarker to cleanly separate the different molecular components.

Cellular morphology and tissue architecture are conventionally identified using colored absorptive transmission stains such as hematoxylin, eosin, Papanicolauo stain (Pap stain), and Masson's trichrome. In contrast, molecular biomarkers can be labeled with either fluorescence probes or transmission stains. Fluorescence labeling generally has the advantage that the spectrum of emission can be quite narrow, and for situations requiring simultaneous tagging of multiple biomarkers, it is possible to choose labels with minimal spectral overlap that are clearly distinguishable—either by eye or with a proper set of interference filters.

Including both fluorescent labels and traditional transmission morphology stains on a single slide makes labeling problematic, though. Hematoxylin and other common stains contain an aluminum "mordant" that quenches fluorescence, making it difficult to obtain adequate signal from fluorescently labeled biomarkers in their presence. In addition, most commonly used transmission stains both absorb and fluoresce strongly themselves throughout the visible wavelength region. Concentrations of these stains vary throughout the sample—a fact that produces two results: a spatially varying decrease in biomarker fluorophore excitation and emission efficiency (due to the stain absorbance), and spatial variations in background fluorescence (due to the stain's own fluorescence). This breaks down the linearity of the spectral addition of each component, and makes linear spectral unmixing invalid.

Thus, multispectral imaging of biomarkers, multiplexed with morphology staining, is significantly more accurate when accomplished using transmission rather than fluorescent biomarker staining.1 Fortunately, many transmission stains are stable and interact minimally with other stains. But transmission stains typically have broad absorbance profiles, so when even just a few of these combine with morphology, results often cannot be interpreted clearly by eye. Multispectral imaging can, however, separate the morphology staining components from those of the various biomarkers even when the stains in question are highly overlapping both spectrally and spatially.

The number of wavelengths needed for accurate spectral imaging depends on the number of stains used, the degree to which they spectrally overlap, and the complexity of the sample being imaged. Figure 1 illustrates this important point. Pap staining of cervical cells is the traditional diagnostic approach to cervical cancer detection. On the far left (a) is a color image of Pap-stained (hematoxylin, eosin, Fast Green, and OG6) cervical cells co-stained for the biomarker p16 with an additional stain, Deep Space Black. There are a total of five stains in the sample, each staining different parts of different types of cells. The unmixed hematoxylin rule image (hematoxylin is the main nuclear stain and of primary importance for morphologic evaluation) for varying number of unmixing wavelengths, N, is shown in the four images to the right (b), with a graph of relative spectral quality shown in (c), and the spectral signatures of the different stains used in the unmixing is shown in (d). For each N, the wavelengths used for the unmixing were chosen to provide the best result possible for this choice of wavelength number. Despite this image containing only five stains, the results using even the best choice of six wavelengths is far from adequate. In fact, achieving the best results requires tens of wavelengths.

The number of wavelengths used in multispectral analysis impacts spectral unmixing image quality: full-color image (a), primary stain images vs. number of wavelengths (b), relative unmixing quality obtained by quantifying the image contrast vs. number of wavelengths (c), and spectra used for the unmixing (d)
FIGURE 1. The number of wavelengths used in multispectral analysis impacts spectral unmixing image quality: full-color image (a), primary stain images vs. number of wavelengths (b), relative unmixing quality obtained by quantifying the image contrast vs. number of wavelengths (c), and spectra used for the unmixing (d).

Real-time multispectral microscopy

Accurately separating the signatures of multiple molecular biomarkers, combined with traditional microscopic imaging of morphology, requires spectral imaging with up to 30 wavelengths. The ideal system combines a high-speed digital camera with a filter that selects wavelengths in order and outpaces the camera frame rate in switching between them. For 30 or more colors, filtering the image with a mechanically rotating filter wheel is not practical. And while liquid-crystal tunable filters (LCTFs) can produce 30 wavelengths, switching takes a large fraction of a second. Acousto-optic tunable filters (AOTFs), however, have electronically programmable spectral flexibility and switching time of <1 ms. We have developed an AOTF-based spectral imaging system that can both image, spectrally unmix the components, and display the results on a computer screen in real time (see http://bcove.me/mnw4b6th).

Figure 2 illustrates the ability of this system to image regular cellular staining and to provide identification using a molecular biomarker. In traditional Pap staining of cervical cells as shown in Fig. 2a, a typical sample contains a wide range of cell shapes and types. Human papilloma virus (HPV) is a well-known biomarker precursor for cancer. We have therefore combined HPV imaging with Pap staining, and Fig. 2b shows how the presence of HPV can be added with a red "false color" overlay to the Pap image-clearly indicating which cells have abnormal DNA.

FIGURE 2. Spectral unmixing can highlight abnormal molecular signatures: Pap-stained cervical cells imaged without a biomarker (a), and the same image is shown with a spectrally imaged HPV biomarker overlay in red (b).

A spectral unmixing system can also help clarify the morphology of cells by allowing separation and accurate intensity adjustment of critical components. Figure 3 demonstrates this on two images of dense clumps of cells. The full image with intense and overlapping stains is opaque, but the control provided by a spectral imaging system allows isolation of the most important hematoxylin stain and reduction of the intensity so interior features can be fully resolved.

Spectral unmixing highlights cellular morphology: (a) and (c) densely stained clumps of cells vs. (b) and (d) unmixed, hematoxylin-only images. Relative intensity of the unmixed components is controlled using a simple software interface (e)
FIGURE 3. Spectral unmixing highlights cellular morphology: (a) and (c) densely stained clumps of cells vs. (b) and (d) unmixed, hematoxylin-only images. Relative intensity of the unmixed components is controlled using a simple software interface (e).

Applications benefit

With the help of AOTFs, images produced by spectral unmixing can be used in conjunction with cell-by-cell biomarker images to produce false-colored overlays with significant additional information to facilitate more accurate interpretation.

By combining spectral information with an image capable of conveying morphology, the potential for this technique is being explored in applications such as cancer detection,2 study of neuronal action potentials,3 blood oxygenation measurement,4 and more. And in combination with other techniques such as surface-enhanced Raman scattering (SERS), it promises even greater outcomes.

REFERENCES

1. E. S. Wachman et al., J. Biomed. Opt., 19, 5, 056006 (2014).

2. R. M. Levenson et al., Expert Opin. Med. Diagn., 2, 1067 (2008).

3. Li Q. et al., Appl. Opt., 52, 17, 3891–3901 (2013).

4. R. D. Shonat et al., Biophys. J., 73, 1223 (1997).

MURRAY REED is chief technology officer; ELLIOT WACHMAN is VP, research and development; AUSTIN DOWD is product manager; and ALEX FONG is senior VP, life sciences and instrumentation for Gooch & Housego; e-mail: afong@goochandhousego.com; www.goochandhousego.com.

Get All the BioOptics World News Delivered to Your Inbox

Subscribe to BioOptics World Magazine or email newsletter today at no cost and receive the latest news and information.

 Subscribe Now
Related Articles

OCT imaging improves percutaneous coronary intervention, study finds

OCT can visualize coronary arteries in patients undergoing percutaneous coronary intervention, leading to improved outcomes.

New bioimaging technique offers clear view of nervous system

Scientists at Ludwig-Maximilians University have developed a technique for turning the body of a deceased rodent entirely transparent, revealing the central nervous system in unprecedented clarity....

Fluorescent jellyfish proteins light up unconventional laser

Safer lasers to map your cells could soon be in the offing -- all thanks to the humble jellyfish. Conventional lasers, like the pointer you might use to entertain your cat, produce light by emittin...

Microscope detects one million-plus biomarkers for sepsis in 30 minutes

A microscope has the potential to simultaneously detect more than one million biomarkers for sepsis at the point of care.

BLOGS

Neuro15 exhibitors meet exacting demands: Part 2

Increasingly, neuroscientists are working with researchers in disciplines such as chemistry and p...

Why be free?

A successful career contributed to keeping OpticalRayTracer—an optical design software program—fr...

LASER Munich 2015 is bio-bent

LASER World of Photonics 2015 included the European Conferences on Biomedical Optics among its si...

White Papers

Understanding Optical Filters

Optical filters can be used to attenuate or enhance an image, transmit or reflect specific wavele...

How can I find the right digital camera for my microscopy application?

Nowadays, image processing is found in a wide range of optical microscopy applications. Examples ...

CONNECT WITH US

            

Twitter- BioOptics World

Copyright © 2007-2016. PennWell Corporation, Tulsa, OK. All Rights Reserved.PRIVACY POLICY | TERMS AND CONDITIONS