OPTICAL COHERENCE TOMOGRAPHY/LIGHT SOURCES: Improved OCT imaging with VCSEL technology

By Benjamin Potsaid, James Jiang, Vijaysekhar Jayaraman, Alex Cable, and James G. Fujimoto

Light sources based on vertical cavity surface-emitting laser (VCSEL) technology enable optical coherence tomography (OCT) to achieve long imaging range, high speed, and flexibility—and promise new applications in biomedicine as well as other realms.

Optical coherence tomography (OCT) is commonly thought of as a millimeter-range imaging technology. With the recent introduction of new OCT light sources based on vertical-cavity surface-emitting laser (VCSEL) technology, by researchers at Thorlabs, Praevium Research, and the Massachusetts Institute of Technology (MIT), this understanding may be about to change.1 VCSEL-based OCT imaging achieves a combination of long imaging range, high speed, and flexibility in operation that potentially opens the door for new applications in biomedicine—as well as other areas.

OCT imaging works like an optical version of ultrasound but instead of sound, infrared (IR) light is used to probe and visualize the structure of tissue and organs. While ultrasound produces image resolution on the scale of a tenth of a millimeter, OCT can achieve orders-of-magnitude finer resolution on the micrometer scale.

In OCT, the axial resolution is related to the bandwidth of light used in the interrogation: The wider the span of wavelengths in the light, the finer the axial resolution. Using a technique called swept-source OCT, the microelectromechanical systems (MEMS)-based tunable VCSEL developed by the researchers achieves fine resolution by repetitively sweeping the output wavelength of light over a broad spectrum in time. It is the manner in which the light is swept that differentiates the VCSEL from previous swept-source OCT technologies.

VCSEL structure and operation

The VCSEL is fabricated using semiconductor processing techniques (see Fig. 1). A laser cavity is formed by locating a gain material between two mirrors, one of which is stationary and the other of which is suspended by a flexible structure. The mirrors form a Fabry-Perot filter such that the wavelength of tuned emission is proportional to the distance separating them. With the application of voltage, an electrostatic MEMS actuator pulls the top mirror down, thereby reducing the cavity length and tuning a shorter wavelength of emission. It is the monolithic design and micron-scale cavity length that enable the VCSEL to produce true single-mode and mode-hop free tuning.

The VCSEL is fabricated using semiconductor processing techniques
FIGURE 1. (a) The VCSEL is fabricated using semiconductor processing techniques (b) on wafers yielding many (c) single VCSEL devices. (d) Traditional OCT swept-source technologies tune many laser modes under the tunable filter envelope. (e) The VCSEL differs from previous technologies in that it tunes a single longitudinal mode. (F) A comparison of OCT technologies shows that the VCSEL does not lose appreciable imaging sensitivity over a 12 mm imaging range while previous technologies show significant sensitivity loss over only a few millimeters. (g) The VCSEL can be operated to achieve a 1.5 m OCT imaging range at high imaging speeds, much longer than previous OCT technologies.

Previously demonstrated OCT swept light sources consisted of a relatively long laser external cavity ranging from a few centimeters to a few meters in length. Tuning was accomplished through a tunable intracavity filter or tunable wavelength selective mirror, which necessarily selected a cluster of modes and provided multimode tuned operation. Shortening the cavity in these designs ends up selecting fewer modes, but introduces discontinuous mode-hopped operation. The VCSEL provides a new operating regime in which a few-microns-long Fabry-Perot cavity comprises the entire laser cavity, pushing the free spectral range (FSR) beyond 100 nm and enabling mode-hop-free single-mode tuning over this entire FSR. The result is that the VCSEL has a coherence length much longer than has ever been seen in OCT swept light source technologies, which enables the unique long OCT imaging range of the VCSEL.2

Long-range imaging

Because current generation OCT technologies are limited to a few millimeters of range, ophthalmic clinics must use three separate OCT instruments for 1) imaging the retina, 2) imaging the anterior eye, and 3) performing axial eye length measurements. A recent paper demonstrates the unique long-range OCT imaging capability of the VCSEL, which enables, for the first time, simultaneous imaging of the anterior eye and measurement of the full length of the eye from the cornea to the retina in a single OCT acquisition (see Fig. 2).3 This constitutes a first step towards an integrated three-in-one ophthalmic instrument able to save cost and space, and streamline patient flow in the clinic.

OCT imaging with a 1060 nm VCSEL
FIGURE 2. OCT imaging with a 1060 nm VCSEL produces (a) a 3D rendering of the anterior eye with an axial eye length measurement to the retina. (b) Non-medical applications demonstrate its broad applicability: (c) the VCSEL enabled imaging of an optical post holder and the OCT volumetric rendering shows depth measurement of deep bore hole. (d) Likewise, an experiment involving a light bulb produced an (e) OCT volumetric rendering that demonstrated the ability of VCSEL technology to produce long range images nondestructively.

This capability to image over longer ranges is promising for many non-medical applications, too. Recent demonstrations include non-destructive imaging inside a light bulb and 15.2 cm imaging of an optical post with precise depth measurement inside a high-aspect-ratio bore hole. Fast 3D OCT imaging of large objects such as these was inconceivable with previous OCT technologies because of limited depth range. The ability to rapidly and non-destructively image in 3D and surface profile over tens of centimeters at high resolution promises to aid in manufacturing processes and quality assurance.

High-speed capture

The current generation of commercial OCT imaging systems operates at between 25 and 100 kHz axial scans rate. The short laser cavity and fast MEMS actuator in the VCSEL enable imaging at speeds up to 1.2 MHz axial scans rate—about 10–50x faster than current technology.

Applications that benefit from high-speed imaging include those involving movement or the need to image a large area. For instance, imaging of the human retina at a 1.2 MHz axial scan rate enables comprehensive coverage of the retina with high lateral pixel density (see Fig. 3). A retinal scan obtained at a 400 kHz axial scan rate can be motion-corrected and averaged from four volumetric acquisitions. For cancer imaging and diagnostics, the high speeds are critical for comprehensive coverage of large areas and to achieve high-lateral-resolution imaging. Using a miniature endoscopic probe, MIT researchers and clinical collaborators produced in vivo images of a rabbit stomach using a 1 MHz axial scan rate as a translational step towards human endoscopic demonstrations.4

(a) an OCT fundus image produced at a 1.2 MHz axial scan rate. This is an example of a large data set of in vivo human retina obtained with a 1050 nm VCSEL. (b) A 400 kHz axial-scan-rate 3D volume of in vivo human retina obtained with a 1050 nm VCSEL was motion-corrected and averaged from four volumes. (c) And a 1 MHz axial-scan-rate 3D volume of in vivo rabbit stomach was obtained with a miniature endoscopic probe with a 1310 nm VCSEL
FIGURE 3. (a) an OCT fundus image produced at a 1.2 MHz axial scan rate. This is an example of a large data set of in vivo human retina obtained with a 1050 nm VCSEL. (b) A 400 kHz axial-scan-rate 3D volume of in vivo human retina obtained with a 1050 nm VCSEL was motion-corrected and averaged from four volumes. (c) And a 1 MHz axial-scan-rate 3D volume of in vivo rabbit stomach was obtained with a miniature endoscopic probe with a 1310 nm VCSEL.

Precise dynamics measurement

Doppler OCT allows the velocity of moving particles in a sample to be measured in the direction of the incident light beam. Precise measurement of the dynamics of the blood flow in the human eye may improve early detection of eye disease. In the eye, the blood enters and exits through the optic nerve head. The high imaging speeds and good phase stability enabled by the VCSEL allow for measurement of rapid blood flow in the optical nerve head (see Fig. 4). Further, high speeds facilitate the collection of multiple repeated 3D volumes to generate a comprehensive time history of blood flow into and out of the eye.5 Plotting the data from each volume, it is possible to observe pulsatility within the eye.

 A comparison of OCT imaging approaches shows (a) an OCT intensity image and a corresponding OCT Doppler image of the optic nerve head in the retina obtained at a 400 kHz axial scan rate. (b) A representative 3D volume of the optic nerve head from a rapidly acquired sequence of OCT volumes enables quantitative measurement of blood flow. (c) Analysis of multiple repeated 3D Doppler volumes of the optic nerve head enabled plotting of pulsatile blood flow into the eye
FIGURE 4. A comparison of OCT imaging approaches shows (a) an OCT intensity image and a corresponding OCT Doppler image of the optic nerve head in the retina obtained at a 400 kHz axial scan rate. (b) A representative 3D volume of the optic nerve head from a rapidly acquired sequence of OCT volumes enables quantitative measurement of blood flow. (c) Analysis of multiple repeated 3D Doppler volumes of the optic nerve head enabled plotting of pulsatile blood flow into the eye.

Phase-based OCT methods such as Doppler in general are much more sensitive to movement than intensity methods, and have been used to measure picometer displacements and micro-resonances in structures. The VCSEL may be beneficial for measuring dynamic phenomena, such as shock waves traveling through materials or dynamic movements of micro and macro structures.

While MIT and Praevium are in the process of deploying VCSEL-based prototypes to clinics with support from the National Institutes of Health Cancer and Eye Institutes (NCI grant R44CA101067 and NEI grant 1R44EY022864-01), Thorlabs is commercializing the MEMS-VCSEL technology to be offered as a complete OCT imaging system, as a benchtop laser, or as a swept source for OEM system integration.

REFERENCES

1. V. Jayaraman et al., Conference on Lasers and Electro-Optics (CLEO), PDPB2 (2011).
2. B. Potsaid et al., "MEMS tunable VCSEL light source for ultrahigh speed 60 kHz-1 MHz axial scan rate and long range centimeter class OCT imaging," Proc. SPIE, 8213, 82130M-8 (2012).
3. I. Grulkowski et al., Opt. Lett., 38, 673–675 (2013).
4. T-H Tsai et al., "Ultrahigh speed endoscopic optical coherence tomography using micro-motor imaging catheter and VCSEL technology" Proc. SPIE, 8571, 85710N (2013).
5. W. Choi et al., Opt. Lett., 38, 338–340 (2013).

Benjamin Potsaid, PhD, is a research scientist at Thorlabs (Newton, NJ) and a visiting scientist at the Research Laboratory of Electronics at MIT (Cambridge, MA); James Jiang, PhD, is the head of OCT research at Thorlabs; Vijaysekhar Jayaraman, PhD, is founder of Praevium Research (Santa Barbara, CA); Alex Cable is founder and president of Thorlabs; and James G. Fujimoto, PhD, is professor of electrical engineering at MIT. Contact Dr. Potsaid at ben.potsaid@gmail.com.

Get All the BioOptics World News Delivered to Your Inbox

Subscribe to BioOptics World Magazine or email newsletter today at no cost and receive the latest news and information.

 Subscribe Now
Related Articles

Eye test that pairs two in vivo imaging methods may detect Parkinson's earlier

A low-cost, noninvasive eye test pairs two in vivo imaging methods to help detect Parkinson's before clinical symptoms appear.

Lightweight handheld probe for OCT provides insight into children's retinas

A handheld device is capable of capturing OCT images of a retina with cellular resolution in infants and toddlers.

Optical brain imaging noninvasively measures small perfusion changes caused by visual stimulation

An optical brain imaging system can track very small, focal changes in cortical perfusion resulting from visual stimulation.

Optical Coherence Tomography: Beyond better clinical care: OCT's economic impact

The optical coherence tomography (OCT) industry has grown dramatically in its first 25 years.

BLOGS

Neuro15 exhibitors meet exacting demands: Part 2

Increasingly, neuroscientists are working with researchers in disciplines such as chemistry and p...

Why be free?

A successful career contributed to keeping OpticalRayTracer—an optical design software program—fr...

LASER Munich 2015 is bio-bent

LASER World of Photonics 2015 included the European Conferences on Biomedical Optics among its si...

White Papers

Understanding Optical Filters

Optical filters can be used to attenuate or enhance an image, transmit or reflect specific wavele...

How can I find the right digital camera for my microscopy application?

Nowadays, image processing is found in a wide range of optical microscopy applications. Examples ...

CONNECT WITH US

            

Twitter- BioOptics World