BIOPHOTONICS: Quasi-CW ultrasound adds mechanical information to optical tomography

By introducing a quasi-continuous-wave (quasi-CW) ultrasound modulation scheme into an ultrasound optical-tomography (UOT) scanner, researchers at Texas A&M University (College Station, TX) have enabled collection of mechanical information from a tissue sample in addition to the optical information normally supplied by UOT.

Developed over a period of years through the cumulative contributions of numerous research groups, UOT extends the tissue penetration depth of optical tomography from millimeters to practically useful distances on the order of centimeters.

After traveling about 200 µm in dense optical tissue, photons tend to lose the memory of their initial direction, phase, and polarization due to multiple scattering events. Interaction with an ultrasound signal, however, can modulate the optical path by varying both the refractive index of the transmission medium and the periodic motion of the physical particles that scatter passing light, which leads to an optical-frequency shift-or “photon tagging”-of light that has passed through the ultrasound signal. Acousto-optic imaging is then enabled by measuring the interference pattern between photons from the same coherent source that have and have not interacted with the ultrasound signal.

Noise introduced into the system by factors such as light speckle, as well as Brownian and other motions of living tissues, tends to limit the potential spatial as well as temporal resolution obtainable from the relatively small signals obtained through UOT. This problem was addressed about three years ago by the introduction of a photorefractive crystal (PRC) to compensate for the noise effects using self-adaptive wavefront holography.1

The Texas A&M researchers have taken this work a step further by applying the ultrasound modulation in the form of a 1-ms-long focused burst. In addition to improving the signal-to-noise ratio of the PRC-based UOT system, this quasi-CW ultrasound modulation also enabled detection of the acoustic-radiation force, which provides mechanical information about the sample.2

The researchers split 532 nm laser light into two paths, one for probing the sample in tandem with the ultrasound signal and the other for a reference signal to pump two wave mixing in the bismuth silicon oxide PRC (see figure). An alternating electric field (1 kHz, 4 kV) enhanced gain in the PRC, and pump-light intensity was maintained at approximately 10 mW/cm2. Measured crystal response time was on the order of 100 ms, and logarithmic improvement in UOT signal strength was observed with increases in ultrasound burst length. The ultrasound source was a focusing ultrasonic transducer with a central frequency of 1 MHz, focal length of 37.5 mm, focal-zone length of 23 mm, and focal-spot diameter of 2.2 mm. The peak acoustic pressure at the focus was 1.5 MPa and the duty cycle was 100 Hz.


Mechanical tissue data was collected using a photorefractive-crystal-based ultrasound optical-tomography system by modulating the ultrasound signal in a quasi-CW fashion.
Click here to enlarge image

The researchers demonstrated UOT imaging based on optical and mechanical properties by imaging segments of a nude mouse tail in a gelatin sample. The UOT image acquired 0.1 ms after the onset of the ultrasound provided optical absorption contrast. Mechanical characteristics of the sample, such as the acoustic impedance mismatch, which occur on a millisecond time scale, were obtained from time-evolution curves of the UOT signal. Time gating the optical signal also allowed decoupling of optical and mechanical information.

Hassaun A. Jones-Bey

REFERENCES

1. F. Ramaz et al., Optics Express 12(22) 5469 (Nov. 1, 2004.

2. X. Xu et al., Optics Letters 32(6) 656 (March 15, 2007).

Get All the BioOptics World News Delivered to Your Inbox

Subscribe to BioOptics World Magazine or email newsletter today at no cost and receive the latest news and information.

 Subscribe Now
Related Articles

NanoScope receives grant for exploring ultrafast laser-based therapeutic gene delivery

A highly photosensitive opsin allows stimulation of retinal cells for restoring vision in patients with retinitis pigmentosa.

Blood becomes a laser emitter for drug testing, cancer treatment

Combining laser light with an FDA-approved green fluorescent dye can monitor cell structure and activity at the molecular level.

Microscopy system harnesses virtual reality technology for image-guided neurosurgery

A microscope image injection system overlays virtual reality imaging onto the brain when viewed through the eyepiece during surgery.

Wearable light therapy device for hair loss receives regulatory approval in Brazil

The iGrow hair growth system, a wearable low-level light therapy device for treating hair loss, is cleared for use in Brazil.

BLOGS

Neuro15 exhibitors meet exacting demands: Part 2

Increasingly, neuroscientists are working with researchers in disciplines such as chemistry and p...

Why be free?

A successful career contributed to keeping OpticalRayTracer—an optical design software program—fr...

LASER Munich 2015 is bio-bent

LASER World of Photonics 2015 included the European Conferences on Biomedical Optics among its si...

White Papers

Understanding Optical Filters

Optical filters can be used to attenuate or enhance an image, transmit or reflect specific wavele...

How can I find the right digital camera for my microscopy application?

Nowadays, image processing is found in a wide range of optical microscopy applications. Examples ...

CONNECT WITH US

            

Twitter- BioOptics World

Copyright © 2007-2016. PennWell Corporation, Tulsa, OK. All Rights Reserved.PRIVACY POLICY | TERMS AND CONDITIONS